114
1. 2.
From the blank screen, press and hold
d + s for three seconds.
You hear three beeps in RUN or STOP
mode when you press the keys and again
after three seconds to conrm that the
Key Lock is unlocked.
The Key Lock remains unlocked during
the programming of the pump.
If no key is pressed for 10 seconds after
your pump has returned to the RUN or
STOP screen, all four keys are automati-
cally locked.
Procedure for unlocking the Key Lock
If you press any key while the screen is blank, the Key Lock
symbol indicates that the keys are locked.
Note The Key Lock remains unlocked during fast bolus delivery
of Standard or Multiwave Bolus.