Table 202: Electrostatic discharge (ESD) and power frequency magnetic elds (all devices)
Guidance and manufacturer’s declaration – electromagnetic immunity
The Accu-Chek Spirit Combo insulin pump is intended for use in the electromagnetic environment specied below. The customer or
the user of Accu-Chek Spirit Combo insulin pump should assure that it is used in such an environment.
Immunity test standard IEC 60601 test level * Compliance level Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If oors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
Not applicable
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to earth
Not applicable