5.10.3 Locking basal rate proles
Your Accu-Chek Spirit Combo insulin pump allows you to adapt
your hourly basal rate prole as needed. You can then lock
your basal rate proles to provide additional protection against
accidentally changing a basal rate during normal operation. If
the basal rate lock is set to ON, programming of the basal rate
proles 1, 2, 3, 4 and 5 is blocked.
If you try to program a basal rate while the BASAL RATE LOCK is
turned ON, the key symbol in the BASAL RATE TOTAL screen
indicates that basal rate programming is locked.
The a and s functions are blocked.
However, you can still change between the different basal rate
proles 1 to 5.
Procedure for locking the basal rate proles
1. 2.
Press d to move to the THERAPY
SETTINGS screen.
Press f to select.
Press d to move to the BASAL RATE
LOCK screen.
Press a or s to turn the basal rate
lock ON or OFF.
Press f to save and exit.