62
Your pump has been tested against, and complies with, the
regulations on electromagnetic interference. Security systems in
airports and anti-theft monitoring devices, such as those in
department stores, should not normally affect the functioning of
your pump.
Because there are so many devices, such as cellular phones,
that emit electromagnetic radiation, the effect of such devices on
your pump cannot be ruled out. It is therefore recommended to
keep your pump at least 4 inches (10 cm) away from these
devices while they are active.
Do not use your pump in hyperbaric chambers or in hazardous
areas, such as areas where ammable gases or vapours could
be present. This could cause an explosion.
Always stop and remove your pump before you enter these
areas. If you have questions, contact your Accu-Chek Customer
Care (see inside back cover).
Your pump is designed to work in normal barometric conditions
(from 500 to 1060 mbar).
3.1.2 Sport and exercise
Exercise is an important part of diabetes management. Your
Accu-Chek Spirit Combo insulin pump is designed to enable you
to safely perform a range of sporting activities while wearing
your pump. It is important that you protect your pump before you
begin exercise or sporting activities. Do not wear your pump
for sports that involve body contact such as boxing, football or
hockey, since rough contact could damage your pump.
Consult our sterile product brochures and accessories catalogue
for more information about carrying systems or contact your
Accu-Chek Customer Care (see inside back cover) for more
information.