Error E6: MECHANICAL ERROR
A mechanical Error has been discovered during a routine
automatic system check and insulin delivery has stopped.
j Press f twice to snooze and to conrm the Error.
j Remove or disconnect the infusion set from your infusion site.
j Remove the cartridge, adapter, infusion set, and battery from
your pump. Insert a new battery.
j It is essential that you rewind the piston rod. Install a new
cartridge and infusion set. (See section 2.7 Inserting the
cartridge). Prime the infusion set (See section 2.8 Priming the
infusion set).
j Reconnect the infusion set at your infusion site.
j Put your pump in RUN mode.
If the Error E6: MECHANICAL ERROR occurs again, use the
alternative therapy plan provided by your doctor or healthcare
team. Contact your Accu-Chek Customer Care (details on inside
back cover) for more assistance.
Error E7: ELECTRONIC ERROR
An electronic Error has been discovered during a routine
automatic system check and insulin delivery has stopped.
Note You cannot turn off and conrm an Error E7: ELECTRONIC
ERROR by pressing f. You must remove the battery
instead.