160
j Press f twice to snooze and to conrm the Error.
j Replace the battery.
j Put your pump in RUN mode.
Note Your pump stores the settings you saved before the
battery power ran out. If your pump is without power for
more than one hour, check that the time and date are
still correct.
See section 1.4 Accessories and disposables on recom-
mended batteries. See section 2.2 Inserting the battery
for instructions on how to change your battery.
Error E3: AUTOMATIC OFF
Automatic Off is a safety feature that stops insulin delivery and
triggers an Error E3: AUTOMATIC OFF if no keys are pressed
within a programmed time period in RUN mode.
j Press f twice to snooze and to conrm the Error.
j Put your pump in RUN mode.