92
4.4.1 Cancelling a Multiwave Bolus
During programming
(the bolus amount
blinks):
There are three ways to cancel a Multiwave Bolus during programming:
j Do not press any key for 20 seconds. The pump returns to the RUN screen.
j Return to the Multiwave Bolus screen by pressing d + a simultaneously.
j Set the total bolus amount to 0.0 units and press f.
No bolus is delivered.
During the start delay
(Multiwave symbol
blinks):
j Press d + a, s or a. The pump beeps and vibrates. The pump
returns to the RUN screen.
No bolus is delivered.
During immediate
bolus delivery
j You can cancel immediate delivery by pressing and holding s or a for
3 seconds until you hear a melody. This cancels the whole bolus (immediate
and Extended Bolus delivery).
Bolus delivery is
interrupted.
During extended
delivery
j You can cancel the extended delivery by putting the pump in STOP mode.
This cancels the Extended Bolus.
Bolus delivery is
interrupted.
Cancellation of the bolus during start delay, immediate bolus
delivery and extended delivery will cause a Warning W8: BOLUS
CANCELLED. Press f twice to conrm and turn off the alert.
See chapter 8 Warning W8: BOLUS CANCELLED.
You can review the actual bolus amount delivered before the
cancellation by looking at the bolus history. See section
5.4.1 Viewing your bolus history.