140
4.
Press d to move to the DELETE DEVICE
screen.
Press f once to delete the pairing and
a second time to conrm.
Then repeat the pairing procedure described in section 6.3 Pairing
your pump with the Meter.
If the pairing is still not successful, contact Accu-Chek Customer
Care (details on inside back cover).
6.4.1 Handling Warnings and Errors
The Accu-Chek Spirit Combo insulin pump and Meter behave in
the following ways when Warnings and Errors occur.
j If both devices are switched on and paired you can enter the
Meter’s Pump menu and snooze and conrm the Warning or
Error through the Meter.
j If the Meter is switched off, and you switch it on in response to
a Warning, Error or Reminder on the pump, the Meter switches
immediately to the Pump menu. All the functions available on
the pump are performed from the Pump menu of the Meter.
Note The above points do not apply in the case of E7:
ELECTRONIC ERROR.