81
4
4.2.2 Cancelling a Quick Bolus
During programming
(the bolus amount
blinks):
Press the s or a key you used to enter the Quick Bolus programming
option. This will reset the bolus amount to 0.0 units. You hear a melody.
If no new bolus is programmed, your pump returns to the RUN screen and
beeps three times and vibrates after ve seconds.
No bolus is delivered.
During conrmation
(when the beeps and
vibrations occur), or
During start delay
(Standard Bolus symbol
blinks):
Press s or a. You hear a melody. Your pump returns to the RUN screen.
A Warning W8: BOLUS CANCELLED occurs.
Press f twice to conrm and turn off the alert. See chapter 8 Warning W8:
BOLUS CANCELLED.
No bolus is delivered.
During bolus delivery
(countdown of bolus
amount):
Press and hold s or a for 3 seconds until you hear a melody. A Warning
W8: BOLUS CANCELLED is displayed.
Press f twice to conrm and turn off the alert. See chapter 8 Warning W8:
BOLUS CANCELLED.
You can view the actual bolus amount delivered before the cancellation by
looking at the bolus history. See section 5.4.1 Viewing your bolus history.
Bolus delivery is
interrupted.
Note Make sure that the cancellation was deliberate, and then program a new bolus, if required.