Loading...
200
Tactile Signals
Dedicated functions of the Accu-Chek Spirit Combo insulin pump
are supported by tactile signals additionally. The tactile signals
are vibrations which are generated by a vibra motor.
Signal Function
SHORT VIBRATION
Duration short vibration =
109.375 ms ±5 %
MEDIUM VIBRATION
Duration medium vibration =
203.125 ms ±5 %
LONG VIBRATION
Duration long vibration = 281.25 ms ±5 %
Signal Function
WARNING AND Reminder
(REPEATING SEQUENCE)
Duration vibration = 1062.5 ms ±5 %
Duration pause = 2125.0 ms ±5 %
(Repeating interval = 3187.5 ms ±5 %)
ERROR (REPEATING SEQUENCE)
Duration vibration = 531.25 ms ±5 %
Duration pause = 1062.5 ms ±5 %
(Repeating interval = 1593.75 ms ±5 %)