57
2
STOP-Warning
The STOP-Warning is triggered when your pump switches from
RUN mode to STOP mode, or when a battery is inserted. The
pump emits two short beeps and vibrates once every minute.
This warns you that insulin delivery has been interrupted.
To turn off the STOP-Warning, press either a or s for three
seconds. You then hear a melody and / or feel a vibration,
conrming that the STOP-Warning has been turned off.
If you press the key too briey, the STOP-Warning will not be
turned off.
Note For your safety, the volume of the STOP-Warning cannot
be changed. It is independent of the programmed beep
volume.
2.12 General safety information
w WARNING
The cartridge and infusion set are sterile products intended
for single use only. Sterility is guaranteed for unopened
packaging up to the indicated expiry date. Do not use sterile
products if the packaging is damaged.
Do not re-use single-use materials because this might lead to
a malfunction of your Accu-Chek Spirit Combo insulin pump,
incorrect insulin delivery and / or infection. Always handle
these items with clean hands. Avoid any contact of the
infusion set and the connecting parts of your insulin pump
with antiseptics, antibiotic creams, soaps, perfumes,
deodorants, body lotions, or any other cosmetics. They may
contaminate these parts.