90
Procedure for programming a Multiwave Bolus
1. 2. 3.
5.
Press d to move to the Multiwave Bolus
screen.
Press f to select. The BOLUS AMOUNT
screen appears.
Press a or s to increase or decrease
the total bolus amount.
Note The rst time you press a,
the bolus duration of the last
Multiwave Bolus and a default
immediate bolus amount appear.
The latter increases or decreases
proportionally as you increase or
decrease the total bolus amount.
Press d to move to the IMMEDIATE
BOLUS screen.
Press a or s to correct and set the
immediate bolus amount.
Press d to move to the BOLUS
DURATION screen.
Press a or s to correct and set the
Extended Bolus duration in increments of
15 minutes for up to 12 hours.
4.