Accu-Chek Spirit Combo - Warning W2: BATTERY LOW; Warning W3: REVIEW TIME

Accu-Chek Spirit Combo
226 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
154
Warning W2: BATTERY LOW
The battery power is low. Be prepared for a battery change as
soon as possible. The battery low symbol appears in the RUN or
STOP screen as a Reminder until you change the battery.
j Press f twice to snooze and to conrm the Warning.
j Replace the battery as soon as possible. Failure to change the
battery within a few hours of receiving the Warning could
result in additional Error messages.
j If necessary, put your pump in RUN mode.
See section 1.4 Accessories and disposables on recommended
batteries. See section 2.2 Inserting the battery for instructions on
how to change the battery.
Warning W3: REVIEW TIME
The pump has been without power for more than one hour. The
pump settings and history data are saved in memory regardless
of power loss. However, it may be necessary to set the time and
date again.
j Press f twice to snooze and to conrm the Warning.
j Review the time and date and correct them, if they are wrong.
j If necessary, put your pump in RUN mode.
See section 2.4 Setting the time and date.