Warning W2: BATTERY LOW
The battery power is low. Be prepared for a battery change as
soon as possible. The battery low symbol appears in the RUN or
STOP screen as a Reminder until you change the battery.
j Press f twice to snooze and to conrm the Warning.
j Replace the battery as soon as possible. Failure to change the
battery within a few hours of receiving the Warning could
result in additional Error messages.
j If necessary, put your pump in RUN mode.
See section 1.4 Accessories and disposables on recommended
batteries. See section 2.2 Inserting the battery for instructions on
how to change the battery.
Warning W3: REVIEW TIME
The pump has been without power for more than one hour. The
pump settings and history data are saved in memory regardless
of power loss. However, it may be necessary to set the time and
date again.
j Press f twice to snooze and to conrm the Warning.
j Review the time and date and correct them, if they are wrong.
j If necessary, put your pump in RUN mode.
See section 2.4 Setting the time and date.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Spirit Combo and is the answer not in the manual?

Accu-Chek Spirit Combo Specifications

General IconGeneral
Device TypeInsulin Pump
DisplayLCD
Water ResistanceIPX8
CommunicationBluetooth
Reservoir Capacity315 units
Bolus CalculatorYes
ManufacturerRoche Diabetes Care
Bolus Dose Increments0.1 units
Battery TypeAAA
Insulin TypeU-100 Insulin
Bolus TypeStandard, Extended
Remote ControlYes
Basal Rate Increments0.01 units/hour
CompatibilityAccu-Chek Combo