166
Error E9: END OF OPERATION (loan pumps only)
Your pump timer has reached zero. The pump changes to STOP
mode and will no longer operate. An Error E9: END OF OPERA-
TION appears on the display. Before this Error occurs, a Warning
W9: LOANTIME WARNING will remind you to replace your pump
soon. If necessary, use the alternative therapy plan provided by
your doctor or healthcare team.
j Press f twice to snooze and to conrm the Error.
j Contact your Accu-Chek Customer Care (details on inside back
cover) to discuss the options for further use of the Accu-Chek
Spirit Combo insulin pump.
Note This setting is country-specic and may not be visible on
your pump. The Warning W9: LOANTIME WARNING and
the Error E9: END OF OPERATION are based on running
time (the total time the pump is in RUN mode) and not on
calendar years.
Error E10: CARTRIDGE ERROR
The CARTRIDGE CHANGE function was not correctly performed.
j Press f twice to snooze and to conrm the Error.
j Remove or disconnect the infusion set from your infusion site.
j Remove the cartridge.
j Reinsert the cartridge. (See section 2.7 Inserting the cartridge)