80
Five seconds after the last press of a,
your pump conrms the total bolus
amount with one beep and vibration for
each bolus increment programmed.
w WARNING
The bolus increment that can be programmed into your
Accu-Chek Spirit Combo insulin pump determines the Quick
Bolus amount using the a and s keys of your insulin
pump.
Incorrect programming of the bolus increment may cause
inappropriate insulin delivery.
See section 5.10.1 Adjusting the bolus increment.
The Standard Bolus symbol blinks for
5 seconds (the bolus delivery start delay).
6.
After the bolus delivery start delay, the
pump beeps three times and vibrates
and then begins to deliver the total bolus
amount programmed. A countdown of
the remaining bolus amount appears on
the display.
5.4.