187
A
Accuracy of ow rate for both
Accu-Chek 3.15 ml Plastic Cartridges
and Accu-Chek Spirit 3.15 ml Cartridge
System
Trumpet curve plotted from data after the end of the
stabilization period
The trumpet curve shows the accuracy of the delivery rate in
relation to the observation period.
percentage Error of ow (%)
20
0
–10
–15
–20
060 120 180 300 660 780 960240 360 840 600 840
10
5
420 480 720 900
–5
15
8.17
–3.17
–2.58
–2.18
0.18
0.22
0.82
3.81
–1.97
0.81
1.79
1.02
A= –1.38
observation window duration (min)
minimum measured variation
maximum measured variation
overall percentage Error
* The measurements were made at a medium basal rate of 1.0 U / h according
to IEC 60601-2-24:1998 with an Accu-Chek Tender infusion
set PC 17 / 110 at room temperature.
The maximum deviation of the delivered amount (overall mean
percentage ow error) for U100 insulin is ≤ ± 5 %*, where your
Accu-Chek Spirit Combo insulin pump is located ≤ 1 m (3.3 feet)
above / below the infusion site.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Spirit Combo and is the answer not in the manual?

Accu-Chek Spirit Combo Specifications

General IconGeneral
Device TypeInsulin Pump
DisplayLCD
Water ResistanceIPX8
CommunicationBluetooth
Reservoir Capacity315 units
Bolus CalculatorYes
ManufacturerRoche Diabetes Care
Bolus Dose Increments0.1 units
Battery TypeAAA
Insulin TypeU-100 Insulin
Bolus TypeStandard, Extended
Remote ControlYes
Basal Rate Increments0.01 units/hour
CompatibilityAccu-Chek Combo

Related product manuals