197
E
Sound Function
+
50 U < CARTRIDGE CONTENT <= 100 U
Frequency: f1 = 2090 Hz ±5 %,
f2 = 2486 Hz ±5 %,
f3 = 2789 Hz ±5 %
Duration: f1 short sound, f2 short sound,
long pause, f3 medium sound
+
100 U < CARTRIDGE CONTENT
Frequency: f1 = 2090 Hz ±5 %,
f2 = 2486 Hz ±5 %,
f3 = 2789 Hz ±5 %
Duration: f1 short sound, f2 short sound,
long pause, f3 long sound
SYSTEM TEST SUCCESSFUL
Frequency: 1043 Hz ±5 %, 1314 Hz ±5 %,
1563 Hz ±5 %,
2090 Hz ±5 %
Duration: 4 x long sound
Sound Function
COPY BASAL RATE
Frequency: 2486 Hz ±5 %, 3131 Hz ±5 %,
2486 Hz ±5 %
Duration: 3 x short sound
Temporary Basal Rate ACTIVE
Frequency: 2789 Hz ±5 %
Duration: Short sound, short pause,
short sound
STOP WARNING
Frequency: 2789 Hz ±5 %
Duration: Long sound, short pause,
long sound

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Spirit Combo and is the answer not in the manual?

Accu-Chek Spirit Combo Specifications

General IconGeneral
Device TypeInsulin Pump
DisplayLCD
Water ResistanceIPX8
CommunicationBluetooth
Reservoir Capacity315 units
Bolus CalculatorYes
ManufacturerRoche Diabetes Care
Bolus Dose Increments0.1 units
Battery TypeAAA
Insulin TypeU-100 Insulin
Bolus TypeStandard, Extended
Remote ControlYes
Basal Rate Increments0.01 units/hour
CompatibilityAccu-Chek Combo