8.
Make sure your infusion set is discon-
nected and the cartridge has been
removed.
Press and hold f for three seconds,
until you hear a melody.
The piston rod begins to return. Do not
touch or otherwise disturb the piston rod
while it is rewinding, as this could result
in damage to your pump.
7.
6.
When the piston rod has returned
completely, the CARTRIDGE VOLUME
screen displays the maximum volume,
315 U.
Press d to move to the CARTRIDGE
CHANGE screen.
Press f to select.
5.
Hold the pump with the adapter facing
the ground. Remove the cartridge, if
necessary. Unscrew carefully. Do not pull
the cartridge out until it is completely
unscrewed.
c CAUTION
Do not change the cartridge if the
cartridge compartment is not completely
dry. When the piston rod is rewinding,
insulin solution could enter your pump
and lead to malfunction. If necessary,
use a soft cloth to dry the compartment.
4.