47
2
The cartridge is correctly inserted when
the end plate of the piston rod is correctly
seated against the cartridge plunger.
Check that the infusion set luer-lock
connector is still correctly connected to
the adapter. Tighten the infusion set
clockwise by hand until it sits tightly in
the adapter.
If the end plate of the piston rod is not
correctly aligned with the cartridge
plunger, remove the cartridge from the
cartridge compartment and repeat steps
9. to 14.
15. 16.
17.
Twist (do not push) the adapter clockwise
until it is correctly seated against the
cartridge compartment.
It is not necessary to apply pressure.
Do not over-tighten the adapter, other-
wise it will be difcult to remove the
cartridge, adapter and cartridge plunger,
and could cause damage to your pump.
13. 14.
Position your insulin pump upright (with
the adapter pointing upwards). Insert the
cartridge into the cartridge compartment.