Accutome A-Scan Plus Connect User’s Guide
Getting Started
Connecting the A-Scan Plus Connect to the Computer
Connect the foot switch to either an unused USB port located with in the
computer or the A-Scan Plus Connect Control Unit.
Connect the receiver of the mouse to an unused USB port located within the
A-Scan Plus Connect Control Unit.
Installation of the Software
Please see the A-Scan Plus Connect Installation Instructions for all Windows
operating systems, which are included in the carrying case.
Connect the A-Scan Plus Connect Control Unit to the computer via USB cable.
A red indicator on the A-Scan Plus Connect Control Unit indicates that the
Control Unit is powered up by the PC but Windows operation system (OS)
is not seeing the Control Unit. If the Windows device driver was installed
correctly, it will change to either yellow or green.
A green indicator on the A-Scan Plus Connect Control Unit indicates that the
Accutome A-Scan Plus Connect is installed correctly and ready to measure
patients.
A yellow indicator on the A-Scan Plus Connect Control Unit indicates that the
A-Scan probe is not connected to the A-scan Plus Connect.
During normal operation, every time the Control Unit is powered up the
indicator will turn red at rst, then turn to green or yellow. A continuous red
(more than 3 seconds) means the device driver is not installed correctly.
Note: Please see table A-1 in the appendix for resolving the problem of a red
and yellow indicator.
The Control Unit will be powered on immediately after being plugged into a
PC and stay on until it is unplugged. To turn it off, just unplug the USB cable
from the PC.
The A-Scan Plus Connect software application is fully functioning even without
the Control Unit. The only limitation is not being able to take new scans.