55
Accutome A-Scan Plus Connect User’s Guide
Personal Preferences
Eye Type Materials and Velocities
The table below list the preset eye types and dene the materials and velocities
for each eye type.
Table 5-2 Eye Types
Eye Type Name Lens Thickness AC VC Extra Lens
Material
Extra
Lens
Thickness
Phakic Phakic Measured Aqueous Vitreous Undened
0
Silicone Filled Phakic Measured Aqueous Silicone
Oil
Undened
0
Dense Cataract Phakic 4.70 Aqueous Vitreous Undened
0
Pseudo PMMA PMMA 0.70 Aqueous Vitreous Undened
0
Pseido Silicone Silicone 1.40 Aqueous Vitreous Undened
0
Pseudo Acrylic Acrylic 0.70 Aqueous Vitreous Undened
0
Aphakic (none) (none) 1532 Undened
0
Phakic+ICL Phakic Measured Aqueous Vitreous
PMMA 0.7
Table 5-3 Lens Materials and Velocities
Material Velocity (m/s)
Phakic 1641
PMMA 2660
Silicone 980
Acrylic 2120
Table 5-4 Anterior/Vitreous Materials and Velocities
Material Velocity (m/s)
Aqueous 1532
Vitreous 1532
Silicone Oil 980