55
Accutome A-Scan Plus Connect Users Guide
Personal Preferences
Eye Type Materials and Velocities
The table below list the preset eye types and dene the materials and velocities
for each eye type.
Table 5-2 Eye Types
Eye Type Name Lens Thickness AC VC Extra Lens
Material
Extra
Lens
Thickness
Phakic Phakic Measured Aqueous Vitreous Undened
0
Silicone Filled Phakic Measured Aqueous Silicone
Oil
Undened
0
Dense Cataract Phakic 4.70 Aqueous Vitreous Undened
0
Pseudo PMMA PMMA 0.70 Aqueous Vitreous Undened
0
Pseido Silicone Silicone 1.40 Aqueous Vitreous Undened
0
Pseudo Acrylic Acrylic 0.70 Aqueous Vitreous Undened
0
Aphakic (none) (none) 1532 Undened
0
Phakic+ICL Phakic Measured Aqueous Vitreous
PMMA 0.7
Table 5-3 Lens Materials and Velocities
Material Velocity (m/s)
Phakic 1641
PMMA 2660
Silicone 980
Acrylic 2120
Table 5-4 Anterior/Vitreous Materials and Velocities
Material Velocity (m/s)
Aqueous 1532
Vitreous 1532
Silicone Oil 980

Table of Contents