11
Impulse equivalent continuous A-weighted sound pressure level (L
AIeq
)
measurement
Impulse equivalent continuous A-weighted sound pressure level (L
AIeq
) is equivalent continuous
sound pressure level with time weighting characteristics, ‘Impulse’.
It can be used only when A characteristics is selected in the default screen.
< Parameter setting >
Measurement is made according to the following procedure
① Frequency weighting key :A (C or Z)
② Display mode key :Leq
③ Time weighting key :Imp
④ Range key :Select a range where the bar graph indicates
approximately 2/3 of the full scale.
【Method of selecting
[Range]
key】
Press
[Range]
key, and choose by cursor keys▲▼, and press
[Range]
key
again to register.
⑤ Measuring time key :1s, 3s, 5s, 10s, 1m, 5m, 10m, 15m, 30m, 1h, 8h, 24h
and***(input to the
[Start/Stop]
key)
The measurement starts with
[Start/Stop]
key.
< Display>
10m 000h00m00s
63.5 dB
40
130
Stp A Imp Leq
Measuring
time
Time weighting:Imp
Display mode
Leq
Rec:Blinks during
recording
Stp: Holding state
Frequency weighting :A (C,Z)
③Time weighting
key
(
F
:Fast,
S
:Slow,
Imp
:Impulse)
①Frequency
weighting key
(A・C, Z)
⑤Measuring
time key
②Display
mode key
④
[Range]
key