REMOTE OPERATION
5-27
Memory sequencing
MSEQ
[not used alone]
:MODE
Select sequencing mode of operation. When a sequence is selected, the user
can step through the sequence using the RCL:UP and RCL:DN commands.
The sequence modes are SEQ1 to SEQ9, and the sequencing can be disabled
with the OFF parameter.
Data type :
Character Program Data
Allowed suffices :
None
Default suffix :
None
Examples:
MSEQ:MODE OFF
MSEQ:MODE SEQ2
:SEQ1...:SEQ9
:START
Set the memory store for the start of the sequencing loop
:STOP
Set the memory store for the end of the sequencing loop
Data type :
Decimal Numeric Program Data
Allowed suffices :
None
Default suffix :
None
Example:
MSEQ:SEQ2:START 50;STOP 70
MSEQ?
Prepares message containing information on the current memory sequencing
mode in the following format:
:MSEQ:MODE <mode>
where: <mode> is character program data indicating the sequence mode
selection
Examples:
:MSEQ:MODE SEQ4
:MSEQ:MODE OFF
MSEQ:SEQ1?
.....
MSEQ:SEQ9?
Prepares message containing information on the start and stop settings of the
given memory sequence in the following format:
:MSEQ:SEQn:START <nr1>;STOP <nr1>
where n is between 1 and 9 inclusive
Example:
:MSEQ:SEQ4:START 120;STOP 155