GENERAL INFORMATION
1-8
Remote control
GPIB: All functions except the supply switch are remotely programmable.
Capabilities: Designed in accordance with IEEE 482.2.
Complies with the following subsets as defined in IEEE Std 488.1: SH1, AH1, T6, TE0,
L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2.
RS-232: All functions except the supply switch are remotely programmable.
Connector: 9-way male D-type.
Bit rate: 300 to 9600 bit/s.
Handshake: Hardware: DTR, RTS, CTS and DSR.
Software: XON and XOFF.
Electrical: Interface to EIA-232-D.
Electromagnetic compatibility
Conforms to the protection requirements of Council Directive 2004/108/EC.
Conforms with the limits specified in the following standards:
IEC/EN 61326-1 : 2006, RF Emission Class B, Immunity Table 1.
Safety
Conforms with the requirements of EEC Council Directive 2006/95/EC and the product
safety standard IEC/EN 61010-1 : 2001 + C1 : 2002 + C2 : 2003 for Class 1 (or 3)
portable equipment, for use in a Pollution Degree 2 environment. The instrument is
designed to operate from an Installation Category 2 or 1+2 supply.
Rated range of use
(over which full specification is met)
Temperature: 0 to 55°C (32 to 131°F).
Humidity: Up to 93% at 40°C (104°F).
Altitude: Up to 3050 m (10 000 ft).
Conditions of storage and
transport
Temperature: −40°C to +71°C (−40 to 160°F).
Humidity: Up to 95% at 40°C (104°F).
Altitude: Up to 4600 m (15 000 ft).
Power requirements
100 to 120 V~, 50 to 400 Hz (limit 90 to 132 V~, 45 to 440 Hz)
210 to 240 V~, 50 to 60 Hz (limit 188 to 264 V~, 45 to 66 Hz
200 VA maximum.
Calibration interval
2 years.
Dimensions and weight
(over projections but excluding front
panel handles)
Height Width Depth Weight
107 mm 419 mm 440 mm <8 kg
4.2 in 16.5 in 17.3 in <17.6 lb
Options
Option 1: No attenuator
Cannot be specified with Option 7 or Option 3.
Omits the internal step attenuator. Specification as standard instrument with the following
exceptions:
RF output range:
From −2 dBm to +15 dBm. When AM is selected the maximum output level reduces
linearly with AM depth to +9 dBm at maximum AM depth.
Pulse modulation: Not available with Option 1.
Output protection: Reverse power protection is not provided.
Option 2: DC operation
Allows for operation from an external DC power source in addition to an AC power source.
Specification as standard instrument with the following additions:
DC supply range: 11 to 32 V.
AC supply requirements 100 to 120 V~, 50 to 400 Hz (limit 90 to 132 V~, 45 to 440 Hz)
210 to 240 V~, 50 to 60 Hz (limit 188 to 264 V~, 45 to 66 Hz
200 VA maximum.
DC consumption: 70 W with Option 3 not fitted. 95 W with Options 3 and 4 fitted.
Rated range of use:
Temperature:
0 to 45°C.