EasyManua.ls Logo

Aeroflex 2023A - Page 29

Aeroflex 2023A
280 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GENERAL INFORMATION
1-9
Option 3: High power
If fast pulse modulation is needed see Option 11.
Specification as standard instrument with the following exceptions:
RF output range: 140 dBm to +25 dBm (output power above +19 dBm is uncalibrated for carrier
frequencies above 1.2 GHz and above +14 dBm above 2.4 GHz).
Maximum output is reduced by 5 dB when pulse modulation is selected and/or by up to
6 dB dependent upon set AM depth.
RF level accuracy (above
+7 dBm):
Over a temperature range 17°C to 27°C (63 to 51°F):
±1 dB 9 kHz to 1.2 GHz <23 dBm
±1.5 dB 9 kHz to 1.2 GHz <25 dBm
±2 dB 1.2 GHz to 2.51 GHz.
Temperature coefficient <±0.02 dB/°C to 1.2 GHz, and <±0.03 dB/°C to 2.51 GHz.
Harmonics: Typically better than –25 dBc for levels 6 dB below the maximum specified output.
Amplitude modulation:
(for RF levels from +7 dBm to
+18 dBm)
Carrier frequency <500 MHz
Standard depth and distortion spec applies
Carrier frequency 500 MHz to 2 GHz
Accuracy:
(from 17°C to 27°C ambient,
temperature coefficient
<±0.02%/°C)
±7.5% of set depth at 1 kHz mod rate
Distortion:
(at 1 kHz mod rate)
<5% up to 80% depth
<2.5% up to 30% depth
Option 4: High stability frequency standard
Replaces the internal TCXO with a high stability OCXO. Specification as standard instrument
with the following exceptions:
Aging rate: Less than ±2.5 in 10
7
per year, less than ±5 in 10
9
per day after 2 months’ continuous
use.
Stability: Better than ±5 in 10
8
over the temperature range 0 to 50°C (32 to 122°F).
Warm-up time: Within 2 in 10
7
of final frequency 10 minutes after switch on at a temperature of 20°C
(68°F).
Option 5: Rear panel connectors
RF output, modulation input and LF output connectors are transferred to the rear panel. The signal
generator’s specification is unaltered.
Option 7: See Annex A.
Option 9: Internal pulse generator
Option 9 adds the ability for the instrument to internally generate pulsed waveforms that modulate
the RF output. Internally-generated pulse waveforms can be free-running, triggered or delay-
triggered.
Modes:
Free-run, externally-triggered, delayed.
Pulse period: 100 ns to 1 s, resolution 50 ns.
Pulse width: 50 ns to 1 s, resolution 50 ns.
Pulse delay: 0 to 1 s, resolution 50 ns.
Active trigger edge: Positive.
External trigger delay: <250 ns.
Jitter: <2 ns.