INSTALLATION
2-10
3. Insulation tests
A 500 V DC test should be applied between the protective earth connection and combined live and
neutral supply connections with the equipment supply switch in the ‘on’ position. It is advisable
to make the live/neutral link on the appliance tester or its connector to avoid the possibility of
returning the equipment to the user with the live and neutral poles linked with an ad-hoc strap.
The test voltage should be applied for 5 seconds before taking the measurement.
Aeroflex employs reinforced insulation in the construction of its products and hence a minimum
pass limit of 7 MΩ should be achieved during this test.
Where a DC power adapter is provided with the equipment, the adapter must pass the 7 MΩ test
limit.
We do not recommend dielectric flash testing during routine safety tests. Most portable appliance
testers use AC for the dielectric strength test, which can cause damage to the supply input filter
capacitors.
4. Rectification
It is recommended that the results of the above tests are recorded and checked during each repeat
test. Significant differences between the previous readings and measured values should be
investigated.
If any failure is detected during the above visual inspection or tests, the equipment should be
disabled and the fault should be rectified by an experienced Service Engineer who is familiar with
the hazards involved in carrying out such repairs.
Safety critical components should only be replaced with equivalent parts, using techniques and
procedures recommended by Aeroflex.
The above information is provided for guidance only. Aeroflex designs and constructs its
products in accordance with International Safety Standards such that in normal use they represent
no hazard to the operator. Aeroflex reserves the right to amend the above information in the
course of its continuing commitment to product safety.
Cleaning
Before commencing any cleaning, switch off the instrument and disconnect it from the supply.
The exterior surface of the case may be cleaned using a soft cloth moistened in water. Do not use
aerosol or liquid solvent cleaners.
Cleaning the LCD window
To prevent damage to the LCD window, care should be taken not to scratch the surface during use
and also when cleaning. The LCD window should be cleaned by wiping a slightly damp, soft,
lint-free cloth gently over the surface
Putting into storage
If the instrument is to be put into storage, ensure that the following conditions are maintained:
Temperature range: −40 to +70°C
Humidity: Less than 93% at 40°C