EasyManua.ls Logo

Aeroflex 2023A - Standard Event Registers

Aeroflex 2023A
280 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE OPERATION
5-39
Standard event registers
This register is defined by IEEE 488.2 and each bit has the meaning shown below:
*ESR?
*ESE
*ESE?
Register
Read/Write
Commands
<esb>
&
&
&
&
&
&
&
&
<pon>
<pon>
<pon>
d
e
<dde>
<dde>
<dde>
d
e
<cme>
<cme>
<cme>
d
e
<rqc>
<rqc>
<rqc>
d
e
<urq>
<urq>
<urq>
d
e
<qye>
<qye>
<qye>
d
e
<exe>
<exe>
<exe>
d
e
<opc>
<opc>
<opc>
d
e
7
7
35
5
1
1
6
6
2
2
4
43
0
0
Transition
Filter #
Condition
Register
Status
Register
C0069
# Positive transition sets status.
Enable
Register
OR
<pon> power on
<urq> user request not implemented in this product
<cme> command error
<exe> execution error
<dde> device-dependent error
<qye> query error
<rqc> request control not implemented in this product
<opc> operation complete set in response to the *OPC command for
synchronization.
<esb> standard event register summary bit

Table of Contents