ACCEPTANCE TESTING
7-34
(2) Connect the test equipment as shown in Fig. 7-17.
(3) On the UUT set:
[CARR FREQ] 30 [kHz]
[RF LEVEL] −4 [dB]
(4) Record the output level measured by the power meter against each of the carrier frequencies
shown in Table 7-45, checking that the resu
lts are within specification.
(5) Set the UUT RF level to +7 dBm and repeat (4) using Table 7-46.
(6)
Set the UUT RF level to +25 dBm and repeat (4) using Table 7-47, decreasing the RF level
t
o +19 dBm when testing at carrier frequencies above 1.2 GHz and further decreasing the RF
level to +14 dBm when testing at carrier frequencies above 2.4 GHz.
ALC linearity
Test procedure
(1) Perform AUTO ZERO and AUTO CAL on the power meter.
(2) Connect the test equipment as shown in Fig. 7-17.
(3) On the UUT set:
[CARR FREQ] 2.5 [MHz]
[RF LEVEL] −4 [dB]
(4) Record the output level measured by the power meter against each of the steps shown in
Table 7-48, checking that the resu
lts are within specification.
(5) Set the UUT carrier frequency to 950 MHz and repeat (4) using Table 7-49.
(6) Set the UUT carrier frequency to 2050 MHz and repeat (4) using Table 7-50.
(7)
Set the UUT carrier frequency to 2510 MHz and repeat (4) using Table 7-51.