REMOTE OPERATION OUTPUT COMMANDS
4-23
OUTPut:MODulation:PM[1]|2[:STATe]
Description: Turns the source feeding the PM1 or PM2 modulator on or off; other active modulators
are not affected. See Fig. 4-1 on page 4-19.
Corresponds to the SOURCE ON/OFF key.
Parameters: <Boolean>
Valid values: OFF | ON | 0 | 1
*RST sets: ON
OUTPut:MODulation:PM[1]|2[:STATe]?
Description: Queries the state of the phase modulation source.
Parameters: None
Response: <Boolean>
Returned values: 0 | 1
OUTPut:MODulation:PULM[:STATe]
Description: Turns the source feeding the pulse modulator on or off; other active modulators are not
affected. See Fig. 4-1 on page 4-19.
Corresponds to the SOURCE ON/OFF key.
Parameters: <Boolean>
Valid values: OFF | ON | 0 | 1
*RST sets: ON
OUTPut:MODulation:PULM[:STATe]?
Description: Queries the state of the pulse modulation source.
Parameters: None
Response: <Boolean>
Returned values: 0 | 1
OUTPut:MODulation:RESet
Description: Equivalent to *RST (page 4-14) for all modulation parameters.