© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-43
Chapter 6: Site Preparation – Power Requirements: Facts about PDUs, MPUs and Power Options
â– In May 2001, the number of power options was
reduced from nine to three and changed as shown
in Table 6-5.
Table 6-5 New Power Options
NEW
Power
Options
Description
1
OLD
Power
Options
Description
(line-to-neutral / line-to-line)
3PD 200–240 V 3-phase Delta 0ED 200 V 3-phase Delta
05C 220 V 3-phase Delta
05H 230 V 3-phase Delta
0E6 240 V 3-phase Delta
3PY 208–220 V 3-phase Wye AWW 120 / 208 V 3-phase Wye with Neutral
05E 127 / 220 V 3-phase Wye with Neutral
3PN 380–415 V 3-phase Wye with Neutral 0E5 220 / 380 V 3-phase Wye with Neutral
05L 230 / 400 V 3-phase Wye with Neutral
0E7 240 / 415 V 3-phase Wye with Neutral
1. Regardless of the power option, the system always operates at 200–240 volts.