© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-97
Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000)
* Three current-carrying conductors
** Four current-carrying conductors
*** Pigtail required (Appendix 10,
Connecting a
Pigtail to the E1135C PDU.)
Table 10-14 3-phase delta (0ED, 05C, 05H, 0E6)
System Type
327X / 79000 317X 307X
Breaker Size in Amps
1
[20] 20 [30] 30 [60] 35
Wire Size mm2 (American Wire Gauge - AWG)
2
[4 (12)] 4(12)** [10(8)] 6(10)* [25 (4)]*** 6(10)
1. Breaker size based on 125 percent of maximum rms current
2. Wire size based on multiconductor cables