© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-99
Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000)
* Three current-carrying conductors
*** Pigtail required (Appendix 10,
Connecting a
Pigtail to the E1135C PDU.)
Table 10-16 Single-Phase Wye with neutral (AWV, AWZ)
System Type
xDSL / POTS Bay 327X / 79000 317X 307X 307X Support Bay
Breaker Size in Amps
1
[20] 20 [35] 30 [60] 30 [60] 35 [30] 20
Wire Size mm
2
(American Wire Gauge - AWG)
2
[4 (12)] 4(12)* [6 (10)] 6(10)* [16(6)] 6(10)*** [16 (6)]*** 10(8)* [6 (10)] 4(12)*
1. Breaker size based on 125 percent of maximum rms current
2. Wire size based on multiconductor cables