8-8 Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
Specifications
Hardware
VSU (Voltage Source Unit) Specifications
Voltage Range, Resolution, and Accuracy
VSU supplemental information
Output resistance typical 0.2 Ω
Capacitive load maximum 10
μF
Slew rate maximum 0.2 V/
μs
Current limit typical 120 mA
Output Noise typical 1 mV rms
Voltage
Range
Resolution
Accuracy
a
a. Accuracy: ± (% of setting value + offset value).
Max.
Output
Current
± 20 V 1 mV
± (0.05 % of setting + 10 mV)
b
b. Specification is applicable under no load current.
100 mA