2-10 Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
Installation
Inspection
To Execute Diagnostics and Self-calibration
You can execute diagnostics and/or self-calibration to verify the operation of the
4155C/4156C.
1. Disconnect all cables from the measurement terminals on the 4155C/4156C rear
panel.
2. Press System key, then CALIB/DIAG softkey.
The SYSTEM: SELF-CALIBRATION/DIAGNOSTICS screen is displayed.
3. If you want to execute both diagnostics and self-calibration, select DIAG
SELFTST ALL softkey.
4. If you want to execute self-calibration only, select CALIB ALL softkey.
5. To stop the diagnostics or self-calibration, select STOP softkey.