Agilent 4155C/4156C User’s Guide Vol.1, Edition 11 8-33
Specifications
General Specifications
General Specifications
• Temperature range
Operating +10 °C to +40 °C (with disk drive operation)
+5 °C to +40 °C (without disk drive operation)
Storage -22 °C to +60 °C
• Humidity range
Operating 20 % to 80 % RH, non-condensing and wet bulb
temperature 29 °C or below (with disk drive operation)
15 % to 80 % RH, non-condensing and wet bulb
temperature 29 °C or below (without disk drive operation)
Storage 5 % to 90 % RH, non-condensing and wet bulb temperature
39 °C or below
• Altitude
Operating 0 to 2,000 m
Storage 0 to 4,600 m
• Power requirement
90 V to 264 V, 47 Hz to 63 Hz
•Maximum VA
4155C/4156C 450 VA (Max)
41501B 350 VA (Max)
• Regulatory Compliance
EMC EN61326-1, AS/NZS CISPR 11
Safety CSA C22.2 No.1010.1-1992
IEC61010-1 / EN61010-1
UL Std No. 61010B-1
Certification CE, CSA(c/us), C-Tick
•Dimensions
4155C/4156C 426 mmW by 235 mmH by 600 mmD
41501B 426 mmW by 190 mmH by 600 mmD
• Weight (approx.)
4155C/4156C 21 kg
41501B 16 kg (option 412. with HPSMU and 2 PGUs)