Agilent 4155C/4156C Users Guide Vol.1, Edition 11 8-25
Specifications
Functions
Stress Force Characteristics
SMU, VSU, and PGU output can be forced for the user specified period.
Stress time 500 μs to 31,536,000 s (365 days)
Resolution 100 μs (500 μs
stress time 10 s)
10 ms (10 s
stress time 31,536,000 s)
Burst pulse count 1 to 65,535 (PGU only)
Trigger The 4155C/4156C outputs a gate trigger while stress channels
are forcing stress.
Knob Sweep
In the knob sweep mode, sweep range is controlled instantaneously with the
front-panel rotary knob. Only the CHANNELS: CHANNEL DEFINITION screen
needs to be defined.
Standby Mode
Measurement units in "Standby" remain programmed to their specified output value
even as other measurement units are reset for the next measurement.
Other Characteristics (measurement control)
Measurement control single, append, repeat, and stop
Stress control stress force, and stop
SMU settings capabilities limited auto-ranging, voltage/current compliance,
power compliance, automatic sweep abort functions,
self-test, and self calibration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 41501B and is the answer not in the manual?

Agilent Technologies 41501B Specifications

General IconGeneral
BrandAgilent Technologies
Model41501B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals