Agilent 4155C/4156C User’s Guide Vol.1, Edition 11 8-21
Specifications
Functions
•VAR1'
Staircase or pulsed sweep synchronized with the VAR1 sweep. Sweep is made
with a user specified ratio and offset value. VAR1' output is calculated as
VAR1' = a × VAR1 + b, where "a" is the user-specified ratio and "b" is the
user-specified offset value.
• CONSTANT
A source unit can be set as a constant voltage or current source depending on the
unit.
•PULSE
One of SMUs can be set as a pulse source.
Pulse width 0.5 ms to 100 ms, 100 μs resolutions.
Pulse period 5 ms to 1 s (≥ pulse width + 4 ms), 100 μs resolutions.
SMU pulse setting accuracy (supplemental information):
Conditions: fixed range measurement except multi channel measurement
Width 0.5 % + 50 μs
Period 0.5 % + 100 μs
Trigger out delay (for pulsed measurements.)
0 to 32.7 ms with 100 μs resolution (< pulse width)