HIGH VOLTAGE SHOCK HAZARD
The Agilent 4155C/4156C/41501B can force dangerous voltages (200 V for
HPSMU, and 100 V for MPSMU) at the force, guard, and sense terminals. To
prevent electric shock hazard, the following safety precautions must be observed
during the use of the Agilent 4155C/4156C/41501B.
Use a three-conductor AC power cable to connect cabinet (if used) and Agilent
4155C/4156C/41501B to an electric ground (safety ground).
If you do not use Agilent 16442 Test Fixture, make sure to connect the INTLK
terminal to a switch that turns off when the shielding box access door is opened.
Confirm periodically that INTLK function works normally.
Before touching the connections of the force, guard, and sense terminals, turn
the Agilent 4155C/4156C/41501B off and discharge any capacitors whenever
possible. If you do not turn the Agilent 4155C/4156C/41501B off, complete
“all” of the following items, regardless of any Agilent 4155C/4156C/41501B
settings.
Terminate measurement by pressing Output key, confim that the Output
status indicator is not lit.
Confirm that the HIGH VOLTAGE indicator is not lit.
Open the shielding box access door (open the INTLK terminal).
Discharge any capacitors if the capacitance is connected to an SMU.
Warn workers in the vicinity of the Agilent 4155C/4156C/41501B about
dangerous conditions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 41501B and is the answer not in the manual?

Agilent Technologies 41501B Specifications

General IconGeneral
BrandAgilent Technologies
Model41501B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals