Early Maintenance Feedback 8
Operation Manual 121
Thresholds
The EMF feature provides two warning thresholds: Service due
and Service warning.
Service Due: When the counter exceeds this number of
injections or days, the Service Due indicator lights and an
entry is made in the Maintenance Log. The Service Due limit
must be larger than the Service warning limit.
Service warning: When the counter exceeds this number of
injections or days, the instrument status shows a
reminder that the component may need maintenance
soon.
Both thresholds are set independently for each counter. You
can enabled one or both, as desired.