EasyManuals Logo

Agilent Technologies 7890B User Manual

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
Early Maintenance Feedback 8
Operation Manual 123
Available Counters
Table 15 lists the most common counters available. The
available counters will vary based on the installed GC
options, consumables, and future updates.
Table 15 Common EMF counters
GC Component Parts with a counter Type Default value
Detectors
FID Collector Number of injections
Jet Number of injections
Ignitor Number of ignition attempts
TCD Switching solenoid On time
Filament on time On time
µECD Insert liner Number of injections
Time since wipe test On time 6 months
NPD Bead Number of injections
Ceramics Number of injections
Collector Number of injections
Bead baseline offset pA Value
Bead baseline voltage Voltage value Ceramic bead: 3.895
Blos bead: 1.045
Bead current integral pA-sec Value
Bead on time On time Ceramic bead: 1200 h
Blos bead: 2400 h
FPD
+
/FPD Ignitor Number of ignition attempts
PMT Number of injections
PMT On time 6 months
Inlets
SSL Gold seal Number of injections 5000
Gold seal Time 90 days
Liner Number of injections 200
Liner Time 30 days
Liner O-ring Number of injections 1000

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals