122 Operation Manual
8 Early Maintenance Feedback
Default Thresholds
Selected counters have default thresholds to use as a
starting point. To view any available information for a
counter:
1 Navigate to the desired counter and press [Enter]. See “To
Enable or Change a Limit for an EMF Counter”.
2 Scroll to the counter’s Service Due entry and press
[Mode/Type]. If available, the default threshold for the
counter displays. Press [Clear] to return to the counter.
If a default limit is not suggested, enter a conservative limit
based on your experience. Use the warning feature to alert
you when service is approaching, then track performance to
determine if the Service Due threshold is too high or too low.
For any EMF counter, you may need to adjust the threshold
values based on the demands of your applications.