EasyManuals Logo

Agilent Technologies 7890B User Manual

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #171 background imageLoading...
Page #171 background image
Configuration 10
Operation Manual 171
To configure an AED transfer line heater
1 Check that a power source for the AED transfer line
heater was assigned. See “To assign a GC power source to
an Aux thermal zone” on page 169.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
Thermal Aux 2, or Thermal Aux 3 depending on where the
AED transfer line heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and
select the AED transfer line, and press [Enter].
To configure an ion trap transfer line heater
1 Check that a power source for the ion trap transfer line
heater was assigned. See “To assign a GC power source to
an Aux thermal zone” on page 169.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
Thermal Aux 2, or Thermal Aux 3 depending on where the ion
trap transfer line heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and
select Ion Trap GC Heated Interface, and press [Enter].

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals