Configuration 10
Operation Manual 171
To configure an AED transfer line heater
1 Check that a power source for the AED transfer line
heater was assigned. See “To assign a GC power source to
an Aux thermal zone” on page 169.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
Thermal Aux 2, or Thermal Aux 3 depending on where the
AED transfer line heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and
select the AED transfer line, and press [Enter].
To configure an ion trap transfer line heater
1 Check that a power source for the ion trap transfer line
heater was assigned. See “To assign a GC power source to
an Aux thermal zone” on page 169.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
Thermal Aux 2, or Thermal Aux 3 depending on where the ion
trap transfer line heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and
select Ion Trap GC Heated Interface, and press [Enter].