EasyManuals Logo

Agilent Technologies 7890B User Manual

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
134 Operation Manual
9 GC-MS Features
To Set Up a Vent Method
A good MS Vent method does the following:
Turns off the MS transfer line heater.
Turns off the inlet heater.
Sets the oven to a low temperature, < 50 °C.
Sets the column flow rate into the MS to as high a flow
as you feel is appropriate and safe. For turbo pumps, set
the flow to 15 mL/min or to the maximum flow rate
possible for the column configuration (note that rates
above 15 mL/min may not provide additional benefits).
For diffusion pumps, typically set the flow to 2 mL/min
(never exceed 4 mL/min).
You must create this method to use the fast venting feature.
To create and store the method:
1 Create the method by making the settings on the GC.
2 When the settings are entered, press [Method].
3 Scroll to MS Vent, then press [Store]. If prompted to
overwrite an existing MS Vent method, press [On/Yes] to
confirm.

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals