EasyManua.ls Logo

Agilent Technologies 7890B - Instrument Configuration Settings

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration 10
Operation Manual 181
Instrument
1 Press [Config]. Scroll to Instrument and press [Enter].
2 Scroll to Serial #. Enter a serial number and press [Enter].
This function can only be done by Agilent service
personnel.
3 Scroll to Auto prep run. Press [On/Yes] to enable Auto prep
run, [Off/No] to disable it. See the Advanced Operation
manual for details.
4 Scroll to Zero Init Data Files.
Press [On/Yes] to enable it. When it is On, the GC
immediately begins to subtract the current detector
output from all future values. This applies only to
digital output, and is useful when a non- Agilent data
system has problems with baseline data that is
non- zero.
Press [Off/No] to disable it. This is appropriate for all
Agilent data systems.
5 Scroll to Require Host Connection. Set On to consider
whether or not the remote host reports Ready as part of
GC readiness.
6 Press [Clear] to return to the Config menu or any other
function to end.

Table of Contents

Other manuals for Agilent Technologies 7890B

Related product manuals