Configuration 10
Operation Manual 181
Instrument
1 Press [Config]. Scroll to Instrument and press [Enter].
2 Scroll to Serial #. Enter a serial number and press [Enter].
This function can only be done by Agilent service
personnel.
3 Scroll to Auto prep run. Press [On/Yes] to enable Auto prep
run, [Off/No] to disable it. See the Advanced Operation
manual for details.
4 Scroll to Zero Init Data Files.
• Press [On/Yes] to enable it. When it is On, the GC
immediately begins to subtract the current detector
output from all future values. This applies only to
digital output, and is useful when a non- Agilent data
system has problems with baseline data that is
non- zero.
• Press [Off/No] to disable it. This is appropriate for all
Agilent data systems.
5 Scroll to Require Host Connection. Set On to consider
whether or not the remote host reports Ready as part of
GC readiness.
6 Press [Clear] to return to the Config menu or any other
function to end.