96 Operation Manual
6 Chromatographic Checkout
If performing a manual injection (with or without a data
system):
a Press [Prep Run] to prepare the inlet for splitless
injection.
b When the GC becomes ready, inject 1 µL of the
checkout sample and press [Start] on the GC.
c The following chromatogram shows typical results for a
new detector with new consumable parts installed.
Sulfur performance
1 Install the sulfur filter and filter spacer.
2 Make the following method parameter changes.
3 Ignite the FPD flame if not lit.
4 Display the signal output and monitor. This output
typically runs between 50 and 60 but can be as high as
Methylparathion
Isooctane
Table 11 Sulfur method parameters (S)
Parameter Value ( mL/min)
H2 flow 50
Air flow 60

Table of Contents