136 Operation Manual
9 GC-MS Features
To Manually Exit the MS Vent State
Normally, exit the MS Vent state when venting is complete
and the MS is ready. When configured with a 5977 series
MSD, the GC will automatically exit the MS Vent state when
the MSD becomes ready again.
1 Press [Aux Det #].
2 Scroll to Clear MS Vent, then press [Enter].
CAUTION
Manually exiting the MS Vent state when the GC and MS are still
connected and the MS is venting or off can damage the MS if you
set inappropriate flows.