GC-MS Features 9
Operation Manual 137
To Use the GC When the MSD is Shut Down
To use the GC while an MS is being repaired or maintained,
do the following:
1 Disable MS communications. Press [Aux Det #], scroll to
MS Communication, then press [Off/No].
2 Scroll to Clear MS Shutdown, then press [Enter].
Be careful to avoid settings that send carrier gas into the
MS, or which raise the temperature of parts which might
cause burns when working on the MS.
If needed, completely uninstall the MS from the GC.