52 Operation Manual
5 Running a Method or a Sequence from the Keypad
To resume a paused sequence
1 Press [Seq Control].
2 Scroll to Resume sequence and press [Enter].
The sequence resumes with the next sample.
To stop a running sequence
1 Press [Seq Control].
2 Scroll to Stop sequence and press [Enter].
The sequence stops at the end of the currently running
subsequence unless [Seq] > Repeat sequence is On. The
sampler tray halts immediately. A stopped sequence can
only be restarted from the beginning.
To resume a stopped sequence
1 Press [Seq Control].
2 Scroll to Resume sequence and press [Enter].
The sequence restarts from the beginning of the sequence.
Aborting a sequence
When a sequence is aborted, it stops immediately without
waiting for the current run to finish.
The following will cause a sequence to abort:
The [Stop] key is pressed.
A sampler error occurs, producing an error message.
The GC detects a configuration mismatch during a method
load.
A running sequence tries to load a method that doesn’t
exist.
The sampler is turned off. You can correct the problem
and then resume the sequence. The aborted sample run
will be repeated.
To resume an aborted sequence
1 Correct the problem. (See "Aborting a sequence".)
2 Press [Seq Control].
3 Scroll to Resume sequence and press [Enter].
The aborted sample run will be repeated.

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals