170 Operation Manual
10 Configuration
6
Scroll to Auxiliary type, press [Mode/Type], scroll to and
select the desired device type, and press [Enter]. Types
may include:
Cryo focus
Cryo trap
AED transfer line
Nickel catalyst
ICMPS argon preheat
ICMPS transfer line
ICPMS injector
Ion Trap GC Heated Interface
G3520 Transfer Line
MSD transfer line
User Configurable
7 When prompted, reboot the GC to implement the changes.
To configure a MSD transfer line heater
1 Check that a power source for the MSD heater was
assigned. See “To assign a GC power source to an Aux
thermal zoneon page 169.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
Thermal Aux 2, or Thermal Aux 3 depending on where the
MSD heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and
select the MSD transfer line, and press [Enter].
To configure a nickel catalyst heater
1 Check that a power source for the Nickel Catalyst heater
was assigned. See “To assign a GC power source to an
Aux thermal zone on page 169.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
Thermal Aux 2, or Thermal Aux 3 depending on where the
Nickel Catalyst heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and
select Nickel catalyst, and press [Enter].

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals