EasyManua.ls Logo

Agilent Technologies 7890B - Enabling or Changing EMF Counter Limits

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
126 Operation Manual
8 Early Maintenance Feedback
To Enable or Change a Limit for an EMF Counter
When using the GC without a data system, enable or change
the limit for a counter as follows:
1 Press [Service Mode].
2 Scroll to Maintenance and press [Enter].
3 Scroll to the desired GC component (front or back inlet,
front or back detector, valves, instrument, and so forth)
and press [Enter] to select it. The GC displays the list of
counters for this component.
4 Scroll to the desired counter.
5 Press [Enter] to select the current counter. The display
will show the Service Due and Service warning entries.
If the Service Due or Service warning line reads a number
or time (number of days, for example), the counter is
enabled.
If the Service Due or Service warning line reads Off, press
[On/Yes] to enable the counter.
The display also shows the date and time when the
counter was last modified.
6 Scroll to each threshold line and enter the desired limit.

Table of Contents

Other manuals for Agilent Technologies 7890B

Related product manuals